FREE TRAINING /

Raising a Defendable Non-Conformity – ISO 13485

Rainié Papadopoulos
3 minutes 33 seconds

Lesson Overview

This video, presented by Rainié Papadopoulos, explains how to raise a defendable nonconformity in compliance with ISO 13485.

Objective & Standards Referenced

  • The goal is to raise clear, easily understood, and defendable nonconformities.
  • Audit results must include identification of relevant processes and conclusions.
  • Key Standards:
    • ISO 13485 – Defines internal audit requirements.
    • ISO 9000 – Provides the definition of a nonconformity.

The Three-Step Approach to Raising a Nonconformity

  • Step 1: Describe the Context & Magnitude
    • Clearly explain what the nonconformity is and how serious the breach is.
  • Step 2: Identify the Breached Clause
    • Provide the numerical reference and exact wording from the standard.
    • Example: A purchasing procedure is generally acceptable but does not consider supplier risk in relation to the medical device.
  • Step 3: Document Objective Evidence
    • Ensure the finding is backed by clear, objective evidence.
    • Example: Supplier selection follows stringent criteria, but there is no documented evidence that device risk was considered.

Final Takeaways

  • A well-documented nonconformity should be objective, clear, and concise.
  • It should remove subjectivity from the audit process.
  • It helps the person responsible for addressing the issue by providing actionable information.